Specifications for the Vaisala Veriteq's Very Low Temperature Logger, the VL-1000-VLT.

Technical Specifications

Size 92x58x28mm
115g
Operating Range -90° C to +35° C. and
0 to 100% RH non-condensing
Measurement Range -90 to 0°C
Accuracy (1 Year) +/- 1°C over -90° C to 0° C
Interfaces RS-232 serial port; half-duplex; 19,200 baud. Contact factory for optional USB interface products.
Mounting Magnetic strips; velcro optional
PC Software Designed for use with Veriteq VeriteqLog Validatable software.
Clock Temperature-compensated over operating range.
Electromagnetic Compatibility FCC Part 15 and CE
Power Source Internal lithium battery with life of 5 years at 1 min. sampling rate. Battery is factory-replaceable.
Memory Type Non-volatile 32K x 8 EEPROM
Memory Size 26,900 12-bit samples
Memory Modes User-selectable: Wrap when memory full (FIFO), or Stop when memory full.
Sample Rate User-selectable (in intervals of 10 seconds) from once per minute to once per day.
Recording span Recording span depends on sample interval selected. Table below details typical sampling rates and the length of time the data logger will retain data in memory before wrapping around or stopping (see Memory Modes).

Sample Interval
Recording Span
1 minute 18.7 days
5 minutes 3.1 months
15 minutes 9.3 months
1 hour 3.1 years

Memory Protection Data retention > 20 years backup without power.
Warranty Two years