Product Specifications:
Multi-Application Data Loggers 1016/1416


General

Size 85 x 59 x 26mm
60g
Operating Range -95°C to +70°C
Interfaces RS-232 serial port; Ethernet; PoE & WiFi interfaces available
Mounting Magnetic strips; 3M Dual Lock™Fasteners; Snap-in connector locks provide secure probe connections
PC Software
  • Vaisala Veriteq Continuous Monitoring System (CMS) Automated monitoring, alarming & reporting
  • OPC server software to add Vaisala Veriteq loggers to any OPC-compatible monitoring system
Clock Accuracy: ±1 min./month at 0°C to +50°C
Electromagnetic Interference Meets FCC Part 15 for digital devices; meets CE requirements for radiated emissions, electrostatic discharge, and radiated susceptibility.
Power Source Internal lithium battery with life of 10 years at 1 minute sampling rate.

Multi-Application Temperature Recorder Accuracy

Sensor Calibrated
Measuring Range
Resolution One year Accuracy
V Range External Probe

-90°C to +50°C

0.01°C at +25°C

±0.35°C over -90°C to +50°C

Memory

Memory Type Non-volatile EEROM
Data Sample Capacity

1016 - 68,600 16-bit samples (2 channels)

1416 - 101,375 16-bit samples (4 channels)

Memory Modes User-selectable: Wrap when memory full (FIFO) or Stop when memory full. User-selectable start time and (for VL series) stop time.
Memory Protection Data retention > 20 years backup without power.
Sampling Rates User-selectable from once every minute to once a day.
Recording Span Recording span depends on sample interval selected and number of channels enabled. Chart below details typical sampling rates and length of time logger will retain data in memory before wrapping around or stopping (see Memory Modes).

Recording Span: 1016 - 22V

Sample Interval
1 Channel
2 Channels
1 minute
1.5 Months
23.8 Days
5 minutes
7.6 Months
3.8 Months
15 minutes
1.9 Years
11.5 Months
1 hour
7.8 Years
3.9 Years

Recording Span: 1416 - 44V

Sample Interval
1 channel
2 channels
3 channels
4 channels
1 minute
2.3 Months
1.1 Months
23.5 Days
17.6 Days
5 minutes
11.3 Months
5.6 Months
3.7 Months
2.8 Months
15 minutes
2.8 Years
1.4 Years
11.3 Months
8.5 Months
1 hour
11.5 Years
5.7 Years
3.8 Years
2.8 Years

Temperature Probes: Stainless or Teflon Tip

Sensor
Operating Range
Color Code
Sensor Tip
Probe Length
Cable Construction
V Range external probe:EPT-23N-XXV

-95°C to +70°C

 

Blue

Stainless Steel Diameter: 3.2 mm length: 38mm

XX denotes length of probe in feet.

3 m and 7.6 m probes available.

2mm Diameter Teflon coated cable
V Range external probe:EPT-22W-XXV
Sealed Teflon Tip Diameter: 3mm Length: 28mm

Temperature Probe Accessories

Thermal Dampening Block EPT-TDB for use in refrigerators and freezers. Simulates a glycol bottle to reduce viewLinc alarms generated by opening and closing a door.