LogTag® TICT
The LogTag TICT can monitor and record statistics against four factory pre-set temperature limits for up to 45 days, displaying a tick mark if environmental conditions remain within limits. If temperatures exceed two high temperature alarm limits or go below two low temperature alarm limits, the LCD shows a REJECT alert, indicating the goods require further checking.
Applications
The LogTag TICT is a low cost disposable electronic temperature indicator for use in monitoring temperature & time statistics in transport applications of up to 45 days duration.
Product Summary
The LogTag TICT is factory pre-configured for one of a range of monitoring profiles. Special profiles are available on request, subject to conditions.
Statistical data can be downloaded from the TICT using the LogTag Indicator interface cradle and freely available LogTag Analyzer operating software if required.
The TICT offers an accurate and cost effective solution for monitoring the cold chain in “last mile” distribution and other applications where full data logging capability is not required.
Product Highlights
- Factory pre-configured to monitor a range of products including standard vaccines.
- Standard 60 minute start delay
- Clear indication of alarm status at all times.
- Monitors for up to 45 days
- Push button Start and Stop.
- Data can be downloaded via an interface cradle and LogTag Analyzer if required.
- Robust and Accurate. Data collected is secure and cannot be manipulated.
- Fits easily into packages and can be mailed via letter rate.
Modes of Operation
Example configuration
- Sampling interval:
- 5 minutes
- Start Delay:
- 60 minutes
- HT1 alarm:
- ≥10 °C instant
- HT2 alarm:
- ≥ 8 °C accumulated for 8 hours
- LT1 alarm:
- ≤2 °Caccumulatedfor 12 hours
- LT2 alarm:
- ≤ 0 °C instant
Product Specifications
Operating Temperature Range
-25 °C to +60 °C
Storage Temperature Range
-25 °C to 65 °C
Ambient humidity range during transport and use
0 to 95 %RH
Resolution
0.1 °C or better in range of -25 °C to +60 °C
Accuracy
- Better than ±0.5 °C for -10 °C to +25 °C
- Better than ±1 °C for -25 °C to -10 °C
- Better than ±1 °C for +25 °C to +60 °C
Capacity
- Minimum & Maximum temperature for each of the 45 days
- 1st activation of each Alarm for each day including Trigger time and duration of each excursion
Memory type
Non volatile
Sampling Interval
Factory set as required. Typically to 5 minute interval.
Logging modes
45 day statistics, temperature/time excursions and Alarm trigger time and duration for each day
Logging Start delay
Factory configuration, typically set to 60 minute interval
Sensor
Precision thermistor
Sensor Reaction time
T90 less than 7 minutes by method detailed in EN12830:1999
Vibration
Withstands vibration specification as detailed in EN12830:1999
Shock
- Withstands shock specification as detailed in EN12830:1999
- Withstand 5 drops 1m to smooth concrete floor without loss of function or calibration
Environmental
IEC 60529: IP 64
Power source
Interne 3V Lithiumbatterie - nicht austauschbar
Battery life:
- Minimum storage life of 18 Months before ‘start’
- Monitoring period: 45 days
- Minimum accessibility (display) period of 6 months after ‘stopped’
Size
71,5 mm (H) x 33,0 mm (B) x 8,6 mm (T) (Volume <21 cm3)
Weight 19 g
Case Material
Polycarbonate
Calibration
Traceable to an ISO/IEC 17025 accredited testing laboratory, to NIST, or to another internationally recognized standards agency
EMV compliance
- EC EMC directives
(EN 61000-6-1:2005 & EN 61000-6-3:2006) - Includes electrostatic discharge as prescribed in EN 61000-4-2
- Complies with FCC Part 15 Subparts A and B
Alarms function & display
- Up to four alarm triggers are indicated on the display
- Display activates when button is pressed
- Display is static allowing photocopying